It was reported that the patient was bleeding from neck, anesthetic had covered with a pressure dressing.Patient's hgb dropped from approximately 147 to approximately 76.At the end of the case, as drapes were removed, it was noted that the bottom chamber of the underbody snuggle warmer on the patients right side was full of blood, was just reaching the level where it was starting to spurt out of the air holes.It was then noted that the neck had been bleeding throughout the procedure-pressure dressing sopping wet, blood dripped down bleeding the neck and into the warming blanket.Intervention required transfusions of blood to the patient to combat the dropped hgb in the or, but was fine post op.Reporter indicated they had did not expect this as a possibility and wanted to make the reporting manufacturer aware, although they do not allege the device worked improperly.
|
Other text: no lot or serial number was provided; therefore, device history record review could not be performed.No product sample was received; therefore, visual and functional testing could not be performed.The reported issue could not be confirmed as no product sample was received for evaluation.If the product is returned, the manufacturer will reopen this complaint for further investigation.E4 is unknown in h10.
|