• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ETHICON ENDO-SURGERY, LLC. LIGAMAX-5MM ENDO CLIP APPLIER; CLIP, IMPLANTABLE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ETHICON ENDO-SURGERY, LLC. LIGAMAX-5MM ENDO CLIP APPLIER; CLIP, IMPLANTABLE Back to Search Results
Model Number EL5ML
Device Problems Failure to Fire (2610); Mechanical Jam (2983)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 06/14/2022
Event Type  malfunction  
Manufacturer Narrative
(b)(4).Batch #: unknown.Report source: health professional.Investigation summary : the product was returned to ethicon for evaluation.Visual inspection and functional testing were conducted on the returned device.Visual analysis of the returned sample determined that the el5ml device was returned with one jaw disengaged from the cam; this condition would not allow the jaws to collapse in order to form the clips and a trocar sleeve inserted.Upon cycling, the instrument was noted to be empty and locked out.The instrument is designed to lockout after all the clips have been fired; therefore a potential cause for the customer reported experience is the firing of all of the clips, as a result, the instrument could no longer be fire due to the activation of the lockout mechanism.In order to evaluate the condition of the internal components of the device, it was disassembled.Upon disassembling, no anomalies were found.The instrument has an orange indicator that appears on the top of the handle as a reference for the user as to the number of clips remaining.A manufacturing record evaluation was performed for the finished device batch number, and no non-conformances were identified.As part of ethicon¿s quality process, all devices are manufactured, inspected, and released to approved specifications.Possible causes for the condition of the jaw disengaged from the cam may be inadvertent force, twisting or pressure being placed on the device jaws, using the jaws of the device as a dissector/retractor or damage to the jaws while entering the trocar.This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon, or its employees that the report constitutes an admission that the product, ethicon, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
 
Event Description
It was reported that during a laparoscopic right hemicolectomy with senhance, the jaws did not close.The clip was not fed into the jaws.The device was attempted to remove from the patient through a trocar, but the jaws did not close, so the device was removed with the trocar.Another device was used to complete the case.There were no adverse consequences to the patient.No further information is available.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LIGAMAX-5MM ENDO CLIP APPLIER
Type of Device
CLIP, IMPLANTABLE
Manufacturer (Section D)
ETHICON ENDO-SURGERY, LLC.
475 calle c
guaynabo 00969
*  00969
Manufacturer (Section G)
ETHICON ENDO-SURGERY, LLC.
475 calle c
guaynabo 00969
*   00969
Manufacturer Contact
orla o'mahony
475 calle c
guaynabo 
*  
329348013
MDR Report Key15204282
MDR Text Key305047156
Report Number3005075853-2022-05202
Device Sequence Number1
Product Code FZP
UDI-Device Identifier10705036001843
UDI-Public10705036001843
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
K050344
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign
Reporter Occupation Other
Type of Report Initial
Report Date 08/10/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received08/10/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberEL5ML
Device Catalogue NumberEL5ML
Device Lot NumberX94A9M
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/06/2022
Date Manufacturer Received07/13/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured02/18/2022
Is the Device Single Use? Yes
Type of Device Usage Initial
Patient Sequence Number1
-
-