Model Number ROB10013
Device Problem Connection Problem (2900)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 07/25/2022
Event Type  malfunction  
Manufacturer Narrative
Internal complaint reference: (b)(4).
 
Event Description
It was reported that, during a cori assisted tka surgery, the real intelligence robotic drill blocked with an internal error message (pfs error).The procedure was completed, with a delay of less than 30 min, using a s+n back-up device.Patient was not harmed.
 
Manufacturer Narrative
Section h5 was updated.Internal complaint reference number: (b)(4).Additional information received by the manufacturer has identified that this event should be re-evaluated for mdr/mir reporting.The new information states that the error that appeared on screen was a connection error from the robotic drill and this failure mode has not caused or contributed to death or serious injury, and has not necessitated medical or surgical intervention to preclude permanent impairment of a body function or permanent damage to a body structure; nor does this failure mode represent a malfunction that could cause or contribute to serious injury if this malfunction was to recur.Therefore, it was determined that this case does not meet the threshold for reporting and is a non-reportable event.If further details are provided confirming the occurrence of a reportable event, our files will be updated accordingly and a further report submitted outlining both the event details and our investigations performed.
 
Event Description
It was reported that, during a cori assisted tka surgery, the real intelligence robotic drill blocked with the following error message: "robotic drill critical error: the robotic drill ha an internal error.".The procedure was completed, with a delay of less than 30 min, using a s+n back-up device.Patient was not harmed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
REAL INTELLIGENCE ROBOTIC DRILL
Type of Device
ORTHOPEDIC STEREOTAXIC INSTRUMENT
Manufacturer (Section D)
BLUE BELT TECHNOLOGIES
2905 northwest blvd ste 40
plymouth MN 55441
Manufacturer (Section G)
BLUE BELT TECHNOLOGIES
2905 northwest blvd ste 40
plymouth MN 55441
Manufacturer Contact
holly topping
7000 west william cannon drive
austin, TX 78735
5123913905
MDR Report Key15204751
MDR Text Key297952269
Report Number3010266064-2022-00559
Device Sequence Number1
Product Code OLO
UDI-Device Identifier00885556757321
UDI-Public00885556757321
Combination Product (y/n)N
Reporter Country CodeBE
PMA/PMN Number
K201022
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,Health Professional,User Facility,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 11/07/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received08/10/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberROB10013
Device Catalogue NumberROB10013
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/09/2022
Is the Reporter a Health Professional? Yes
Date Manufacturer Received07/25/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
Treatment
REAL INTELLIGENCE CORI: ROB10024, SN000313
-
-