Catalog Number 364413
Device Problem Inability to Auto-Fill (1044)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 06/13/2022
Event Type  malfunction  
Manufacturer Narrative
Investigation summary: bd had not received samples or photos for investigation.Therefore, 10 retention samples from bd inventory were evaluated by visual examination and functional testing, each drawn with water, and no issues were observed relating to insufficient flow as all samples met specifications.Based on a review of the device history record for the incident lot, all product specifications and requirements for lot release were met.There were no related quality issues during manufacturing of the product.This complaint is unable to be confirmed for the indicated failure mode insufficient blood flow.Bd was not able to identify a root cause for the indicated failure mode.Our business team regularly reviews the collected data for identification of emerging trends.".
 
Event Description
It was reported when using the bd a-line¿ (syringe w/ needle), the device experienced insufficient blood flow through the device.This event occurred 1869 times.The following information was provided by the initial reporter.The customer stated: the medical device does not have sufficient filling pressure for gas collection, which leads to the use of another additional device to complete the procedure.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BD A-LINE¿ (SYRINGE W/ NEEDLE)
Type of Device
BLOOD SPECIMEN COLLECTION DEVICE
Manufacturer (Section D)
BECTON, DICKINSON AND COMPANY (BD)
belliver way
belliver industrial estate
plymouth
UK 
Manufacturer (Section G)
BECTON, DICKINSON AND COMPANY (BD)
belliver way
belliver industrial estate
plymouth
UK  
Manufacturer Contact
phillip emmert
9450 south state street
sandy, UT 84070
8015296192
MDR Report Key15207618
MDR Text Key305372997
Report Number9617032-2022-00751
Device Sequence Number1
Product Code JKA
Combination Product (y/n)Y
Reporter Country CodeCO
PMA/PMN Number
K022426
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,User Facility
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 08/04/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received08/11/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Expiration Date10/31/2023
Device Catalogue Number364413
Device Lot Number1299181
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received07/19/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured10/26/2021
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age67 YR
Patient SexFemale
-
-