Investigation summary: bd had not received samples or photos for investigation.Therefore, 10 retention samples from bd inventory were evaluated by visual examination and functional testing, each drawn with water, and no issues were observed relating to insufficient flow as all samples met specifications.Based on a review of the device history record for the incident lot, all product specifications and requirements for lot release were met.There were no related quality issues during manufacturing of the product.This complaint is unable to be confirmed for the indicated failure mode insufficient blood flow.Bd was not able to identify a root cause for the indicated failure mode.Our business team regularly reviews the collected data for identification of emerging trends.".
|
It was reported when using the bd a-line¿ (syringe w/ needle), the device experienced insufficient blood flow through the device.This event occurred 1869 times.The following information was provided by the initial reporter.The customer stated: the medical device does not have sufficient filling pressure for gas collection, which leads to the use of another additional device to complete the procedure.
|