A us distributor contacted zoll to report that a patient passed away in the hospital while wearing the lifevest on (b)(6) 2022.Review of the download data, indicates the patient received one appropriate treatment during vf.From 09:19:28 to 09:21:42, the patient was seen in vt from 210-240 bpm.Motion artifact and varying amplitudes prevented the lifevest from treating the patient.At 09:22:22, the patient received the appropriate treatment from the lifevest.The patients rhythm at the time of the treatment was vf.The patient's post-shock rhythm was sinus bradycardia at 20 bpm with pvcs and unconducted p waves.At 09:38:10, the patient received a non-lifevest defibrillation.Rhythm at the time of the defibrillation was svt at 200 bpm.Post shock rhythm was sinus tachycardia at 140 bpm with pvc¿s.The lifevest typically requires 60 seconds of sustained vt to deliver a shock.At 10:04:17, the device was shutdown.Patient was last seen in sinus tachycardia at 100 bpm with motion artifact and electrode lead fall off.The patient passed away on (b)(6) 2022.
|