• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: VDW GMBH WST 6:1 F. VDW.GOLD/SILVER; HANDPIECE, CONTRA- AND RIGHT-ANGLE ATTACHMENT, DENTAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

VDW GMBH WST 6:1 F. VDW.GOLD/SILVER; HANDPIECE, CONTRA- AND RIGHT-ANGLE ATTACHMENT, DENTAL Back to Search Results
Catalog Number V041079000000
Device Problems Detachment of Device or Device Component (2907); Mechanics Altered (2984)
Patient Problem Insufficient Information (4580)
Event Type  malfunction  
Event Description
In this event it is reported that wst 6:1 f.Vdw.Gold/silver won't hold files.Outcome of patient injury is unknown as of this mdr and information has been requested.
 
Manufacturer Narrative
There has been a previous report received with a similar device where this malfunction resulted in a serious injury.Therefore, it must be presumed that recurrence of this malfunction could possibly cause or contribute to a serious injury or require medical or surgical intervention to preclude such.As such, this event is reportable per 21cfr part 803.The device is available for evaluation, though has not been returned as of this report.Evaluation results will be submitted as they become available.
 
Manufacturer Narrative
Inspection found head drive defective.The contra angle is very dirty.Cause traced to maintenance.Multiple unsuccessful attempts were made to obtain the patient outcome.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
WST 6:1 F. VDW.GOLD/SILVER
Type of Device
HANDPIECE, CONTRA- AND RIGHT-ANGLE ATTACHMENT, DENTAL
Manufacturer (Section D)
VDW GMBH
bayerwaldstrasse 15
munich bayern 81737
GM  81737
Manufacturer (Section G)
VDW GMBH
bayerwaldstrasse 15
munich bayern 81737
GM   81737
Manufacturer Contact
hannah seevaratnam
221 west philadelphia st.
york, PA 17401
7178457511
MDR Report Key15210040
MDR Text Key302262830
Report Number9611053-2022-00607
Device Sequence Number1
Product Code EGS
Combination Product (y/n)N
Reporter Country CodePL
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional
Reporter Occupation Dentist
Type of Report Initial,Followup
Report Date 09/09/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Catalogue NumberV041079000000
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Distributor Facility Aware Date08/08/2022
Initial Date Manufacturer Received 08/08/2022
Initial Date FDA Received08/11/2022
Supplement Dates Manufacturer Received08/08/2022
Supplement Dates FDA Received09/09/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Type of Device Usage A
Patient Sequence Number1
-
-