Model Number 284504
Device Problems Overfill (2404); Output Problem (3005)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 07/27/2022
Event Type  malfunction  
Event Description
This is report 1 of 2 for (b)(4).It was reported by a healthcare professional in france that during an unknown surgery on (b)(6) 2022, it was observed that while purging the "green notch" fms fluid management system/inflow tubing (fms duo+ or fms solo) device, the reservoir filled-up completely then the pump went into alarm.According to the report, the pump was changed and the same problem occurred with another set of tubing device from the same lot.There were no adverse patient consequences nor surgical delay reported.No additional information was provided.
 
Manufacturer Narrative
This report is being submitted in pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by depuy mitek or its employees that the report constitutes an admission that the device, depuy mitek, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.Udi: (b)(4).Investigation summarythe complaint device is not being returned, it was discarded by the customer, therefore unavailable for a physical evaluation.A manufacturing record evaluation was performed for the finished device lot number (3003555), and no non-conformance was identified.As part of depuy mitek¿s quality process all devices are manufactured, inspected, and released to approved specifications.With the information provided, and without the complaint device to evaluate, we cannot determine a root cause for the reported failure.Additional complaint information monitoring for potential safety signals is conducted through complaint trending as part of post market surveillance.At this point in time, no corrective action is required, and no further action is warranted.Depuy mitek will continue to track any related complaints within this device family as a means of monitoring the extent with which this complaint is observed in the field.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FMS FLUID MANAGEMENT SYSTEM/INFLOW TUBING (FMS DUO+ OR FMS SOLO)
Type of Device
SURGICAL IRRIGATION TUBING SET, GENERAL-PURPOSE
Manufacturer (Section D)
DEPUY MITEK LLC US
325 paramount drive
raynham MA 02767
Manufacturer (Section G)
DEPUY MITEK LLC US
325 paramount drive
raynham MA 02767
Manufacturer Contact
kate karberg
325 paramount drive
raynham, MA 02767
3035526892
MDR Report Key15212923
MDR Text Key305419538
Report Number1221934-2022-02470
Device Sequence Number1
Product Code HRX
UDI-Device Identifier10886705017393
UDI-Public10886705017393
Combination Product (y/n)N
Reporter Country CodeFR
PMA/PMN Number
K951843
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,User Facility
Reporter Occupation Other Health Care Professional
Type of Report Initial
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date01/30/2024
Device Model Number284504
Device Catalogue Number284504
Device Lot Number3003555
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 07/29/2022
Initial Date FDA Received08/11/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured02/15/2022
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-