• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ABBOTT VASCULAR EMBOSHIELD NAV6 EMBOLIC PROTECTION SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ABBOTT VASCULAR EMBOSHIELD NAV6 EMBOLIC PROTECTION SYSTEM Back to Search Results
Model Number 22438-19
Device Problems Difficult to Remove (1528); Failure to Advance (2524)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 07/08/2022
Event Type  Injury  
Event Description
It was reported that the procedure was to treat a lesion in the right internal carotid artery.The emboshield nav6 embolic protection system (eps) was advanced to it's parking zone and deployed without issue.A non-abbott stent was implanted.The retrieval catheter for the eps filter was attempted to be advanced to retrieve the filter; however, it failed to cross the newly implanted stent, so it was removed from the patient.A lower profile balloon and guide liner were then advanced and used to pull the filter into the tip of the guide liner.After visualizing that the filter was collapsed, the devices were removed.There was no adverse patient effect and no clinically significant delay reported.No additional information was provided.
 
Manufacturer Narrative
The device was not returned for evaluation.The lot history record (lhr) for this product could not be reviewed and a similar complaint query could not be performed because the product was not returned for evaluation and the lot number was not reported.Based on the information provided, the reported difficulties appear to be due to circumstances of the procedure.The difficulty advancing the retrieval catheter and difficulty removing the filter resulting in unexpected medical intervention appears to be due to anatomical conditions and interaction with the newly implanted stent.There is no indication of a product quality issue with respect to the design, manufacture, or labeling of the device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EMBOSHIELD NAV6 EMBOLIC PROTECTION SYSTEM
Type of Device
EMBOLIC PROTECTION SYSTEM
Manufacturer (Section D)
ABBOTT VASCULAR
26531 ynez rd.
temecula CA 92591 4628
Manufacturer (Section G)
ABBOTT VASCULAR, REG # 2024168
26531 ynez road
temecula CA 92591 4628
Manufacturer Contact
lindsey bell
26531 ynez rd.
temecula, CA 92591-4628
9519143996
MDR Report Key15214858
MDR Text Key297758930
Report Number2024168-2022-08705
Device Sequence Number1
Product Code NTE
UDI-Device Identifier08717648137754
UDI-Public08717648137754
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K141678
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial
Report Date 08/11/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number22438-19
Device Catalogue Number22438-19
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 07/20/2022
Initial Date FDA Received08/11/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient Age83 YR
Patient SexMale
-
-