DORNOCH DUO FLUID CART WITH SMOKE EVACUATOR; APPARATUS, SUCTION, WARD USE, PORTABLE, AC-POWERED
|
Back to Search Results |
|
Model Number N/A |
Device Problem
Failure to Power Up (1476)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 07/20/2022 |
Event Type
malfunction
|
Event Description
|
It was reported unit will not power on when plugged into wall.The event timing was not during surgery.No patient involvement or harm.During investigation, it was found that there were signs of burn, fire, char.No harm noted and no smoke was involved.No further event information available at the time of this report.
|
|
Manufacturer Narrative
|
(b)(4).The results of the investigation are as follows: review of the most recent repair record determined the power inlet module fuse ports were blistered with blown fuses.The power inlet module and the power module connectors were replaced and resolved the reported issue.A definitive root cause cannot be determined; the event is confirmed.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
|
|
Search Alerts/Recalls
|
|
|