Model Number 490102
Device Problem Device Misassembled During Manufacturing /Shipping (2912)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Type  malfunction  
Manufacturer Narrative
This report has been identified as b.Braun medical internal report number (b)(4).The investigation is ongoing at this time.A follow up will be submitted when the investigation results become available.
 
Event Description
As reported by the user facility: event 17: that the green clip is on the wrong side and it will not load into the pumps.No injury reported.
 
Manufacturer Narrative
Event 17: this report has been identified as b.Braun medical internal report number (b)(4).Eighteen samples 10 used without packaging and 8 unused in unopened packaging were returned by the facility for evaluation.The 18 samples provided were all visually evaluated and it was noted all 18 samples had an inverted space pump.The reported defect is confirmed.Incidents of this nature are attributed to operator oversight during the assembly of the product.Although our training procedures ensure that all employees are properly trained in their areas of responsibility, an oversight on the part of the operator can attribute to an incident of this nature.As a result of this occurrence, a formal awareness training session was conducted with all applicable personnel involved in the assembly and inspection of this product.The purpose of this training was to review the reported incident and to ensure all personnel understand and comply with the established assembly and inspection processes.Review of the discrepancy management system (dsms) database was performed for the reported lot number and no abnormalities or non-conformances were noted during the in process or final product inspection.We will maintain this report for further references and continue to monitor other reports for similar occurrences.If any additional pertinent information becomes available, a follow up will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INFUSOMAT®
Type of Device
SET, ADMINISTRATION, INTRA
Manufacturer (Section D)
B. BRAUN MEDICAL INC.
901 marcon blvd.
allentown PA 18109
Manufacturer (Section G)
B. BRAUN DOMINICIAN REPULIC INC.
las americas industrial park
km22 autopista las americas
santo domingo
DR  
Manufacturer Contact
jonathan severino
901 marcon blvd.
allentown, PA 18109
4847197287
MDR Report Key15217831
MDR Text Key302838373
Report Number2523676-2022-00413
Device Sequence Number1
Product Code FPA
UDI-Device Identifier04046964186127
UDI-Public(01)04046964186127
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K142036
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 09/23/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received08/12/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number490102
Device Catalogue Number490102
Device Lot Number0061822587
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Date Manufacturer Received07/27/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured02/23/2022
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-