OAKDALE LEVEL 1 HOTLINE FLUID WARMER; WARMER, THERMAL, INFUSION FLUID Back to Search Results
Catalog Number CON-HL-90
Device Problems Overheating of Device (1437); Pumping Problem (3016)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Type  malfunction  
Manufacturer Narrative
A product sample was received and is awaiting evaluation and investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.Udi is unknown, no product information has been provided to date.
 
Event Description
It was reported that the pump was bad/over heating.No patient injury was reported.Additional information was requested; however, no further information was received.
 
Event Description
Additional info received via email 6-sep-2022.The event occurred during preventive maintenance.No patient involvement.
 
Manufacturer Narrative
Manufacturing device history record review was not performed because the results of the complaint investigation do not indicate a problem with the manufacture of the device.No causes or potential causes of the customer's reported problem were found during the review of service and repair records the device was returned back with a cracked tank cover, and a wear and tear damaged front cover.The pump was noisy and running hot confirming, therefore the customer's complaint was confirmed.No action was taken, due to the age and condition of the device.It was deemed beyond economical repair and will be scrapped.A manufacturing device history record review was not performed because the results of the complaint d5 and e4 are unknown.Additional event information received indicating no patient involvement (updated b5).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LEVEL 1 HOTLINE FLUID WARMER
Type of Device
WARMER, THERMAL, INFUSION FLUID
Manufacturer (Section D)
OAKDALE
3350 granada ave n
oakdale MN 55128
Manufacturer (Section G)
NULL
3350 granada ave n
oakdale MN 55128
Manufacturer Contact
jim vegel
6000 nathan lane north
dock 5
minneapolis, MN 55442
MDR Report Key15221135
MDR Text Key302029757
Report Number3012307300-2022-15327
Device Sequence Number1
Product Code LGZ
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K001764
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type User Facility
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 09/12/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received08/12/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberCON-HL-90
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/02/2022
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Date Manufacturer Received08/15/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured07/18/2008
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-