ST. JUDE MEDICAL, INC.(CRM-SYLMAR) ALLURE QUADRA RF CRT-P; IMPLANTABLE PACEMAKER PULSE GENERATOR Back to Search Results
Model Number PM3242
Device Problem Migration or Expulsion of Device (1395)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 07/13/2022
Event Type  Injury  
Event Description
It was reported the patient presented for a pacemaker exchange.Prior to the procedure, it was noted the pacemaker had migrated.The pacemaker was explanted and replaced in the proper anatomic position.The patient was in stable condition.
 
Manufacturer Narrative
The reported event of migration was not confirmed.The pacemaker was returned and elective replacement indicator (eri) for analysis.Longevity and visual analysis were normal with no anomalies found.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ALLURE QUADRA RF CRT-P
Type of Device
IMPLANTABLE PACEMAKER PULSE GENERATOR
Manufacturer (Section D)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer (Section G)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer Contact
sonali arangil
15900 valley view court
sylmar, CA 91342
MDR Report Key15222107
MDR Text Key297856399
Report Number2017865-2022-17502
Device Sequence Number1
Product Code NIK
UDI-Device Identifier05414734507912
UDI-Public05414734507912
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P030035
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 09/01/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date08/31/2016
Device Model NumberPM3242
Device Catalogue NumberPM3242
Device Lot NumberA000002264
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/25/2022
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 07/25/2022
Initial Date FDA Received08/12/2022
Supplement Dates Manufacturer Received08/25/2022
Supplement Dates FDA Received09/01/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured02/24/2015
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
QUARTET; TENDRIL; TENDRIL STS
Patient Outcome(s) Required Intervention;
Patient Age88 YR
Patient SexFemale
-
-