The log file and the flow and pressure measurement module as well as the two corresponding data cables were provided for investigation.Based on the log file it could be verified that the device preformed a synchronized restart of the ventilation unit and display unit on august 01, 2022 whilst ventilation was active.It was found that the restart was caused in specified reaction on a detected adc failure of the flow and pressure measurement module.After successful restart of the device the ventilation continued until the ventilation was set into standby mode by the user.The device posted corresponding alarms to alert the user.The hardware analysis of the flow and pressure measurement module and corresponding data cables could not reveal any deviation.As a safety feature of the device, the software analyses and verifies proper function of the device.In case of a detected failure concerning the flow and pressure measurement module the device would perform a restart to mitigate the issue and post an audible alarm to alert the user.During restart the emergency-breathing valve will open to ambient allowing for spontaneous breathing.The device reacted as specified.The number of similar cases, related to the same root cause, is within the expected range of the respective risk assessment and thus accepted.
|