This event has been recorded by zimmer biomet under (b)(4).Telephone number: (b)(6).Review of the most recent repair record identified the following related repair: tech confirmed the unit was not working due to the cable having contact issues.Additionally, the tech found that the unit's motor speed was unstable.Tech replaced the motor and plug harness assembly.The unit was tested, calibrated, and returned to the customer.Review of the device history record identified no deviations or anomalies during manufacturing related to the reported event.A definitive root cause cannot be determined.The event is confirmed.If any further information is found which would change or alter any conclusions or information, a supplemental report will be filed accordingly.Zimmer biomet will continue to monitor for trends.
|