Devilbiss healthcare was notified by an oxygen supplier of a complaint involving a devilbiss suction unit stating that the unit is "not suctioning properly." there was no report or evidence of illness, injury or medical treatment associated with the complaint.The device was returned for evaluation.Devilbiss evaluated the unit and determined that the primary cause was an umbrella valve being dislodged from the compressor cylinder, manifold, broken/damaged suction gauge.Devilbiss has an active capa associated with cylinder/valve complaints.
|