ST. JUDE MEDICAL, INC.(CRM-SYLMAR) QUARTET LEAD SMALL-S, 86 CM; PERMANENT PACEMAKER ELECTRODE Back to Search Results
Model Number 1456Q/86
Device Problems Break (1069); Failure to Disconnect (2541)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 07/27/2022
Event Type  malfunction  
Event Description
It was reported that the patient presented for a scheduled procedure.During the procedure, the physician implanted the left ventricular lead and the stylet got stuck in the lead while removing and the inner coil got damaged.The lead was removed and replaced.The patient was in stable condition.
 
Manufacturer Narrative
The reported events of stylet could not be removed and lead damaged were confirmed.As received, a complete lead was returned in one piece with the stylet stuck inside the lead.Visual inspection of the lead found that the connector pin and connector cap were pulled out of the connector assembly and the inner coil was stretched consistent with procedural damage and the ptfe stylet coating was bunched up/clogged with the inner coil distal to the connector pin.The cause of the reported events was due to bunched up ptfe coating of the stylet that prevented the removal of the stylet and excessive forces resulted in the connector pin, cap and inner coil to be pulled out of the connector assembly.A review of the device history record (dhr) confirmed that no issues were identified related to this reported event.Abbott is continuing to monitor this issue.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
QUARTET LEAD SMALL-S, 86 CM
Type of Device
PERMANENT PACEMAKER ELECTRODE
Manufacturer (Section D)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer (Section G)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer Contact
sonali arangil
15900 valley view court
sylmar, CA 91342
MDR Report Key15236470
MDR Text Key302876543
Report Number2017865-2022-19506
Device Sequence Number1
Product Code NIK
UDI-Device Identifier05414734510158
UDI-Public05414734510158
Combination Product (y/n)Y
Reporter Country CodeIN
PMA/PMN Number
P030054
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,User Facility,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 09/26/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received08/16/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number1456Q/86
Device Catalogue Number1456Q/86
Device Lot NumberA000116359
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received09/22/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured10/07/2021
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-