• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ABBOTT VASCULAR XIENCE XPEDITION EVEROLIMUS ELUTING CORONARY STENT SYSTEM; DRUG ELUTING CORONARY STENT DELIVERY SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ABBOTT VASCULAR XIENCE XPEDITION EVEROLIMUS ELUTING CORONARY STENT SYSTEM; DRUG ELUTING CORONARY STENT DELIVERY SYSTEM Back to Search Results
Catalog Number 1070300-38
Device Problem Tear, Rip or Hole in Device Packaging (2385)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 04/23/2022
Event Type  malfunction  
Manufacturer Narrative
A visual inspection was performed on the returned device.The reported tear, rip or hole in device packaging could not be tested due to the condition of the returned chipboard box.A review of the lot history record identified no manufacturing nonconformities issued to the reported lot that would have contributed to this event.Additionally, a review of the complaint history identified no similar incidents from this lot.A conclusive cause for the reported tear, rip or hole in device packaging could not be determined.There is no indication of a product quality issue with respect to manufacture, design, or labeling.
 
Event Description
It was reported that upon opening the package, the 3.0x38mm xience xpedition stent delivery system (sds) outer box was tampered and noted to be opened.The sticker on the chipboard box tab was torn.The inner package and device were not opened and remained intact.There was no device use or patient involvement.There was no clinically significant delay in the procedure.Returned device analysis identified that the inner pouch was opened.It was confirmed that the account was aware of the inner pouch being opened upon using the device.No additional information was provided.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
XIENCE XPEDITION EVEROLIMUS ELUTING CORONARY STENT SYSTEM
Type of Device
DRUG ELUTING CORONARY STENT DELIVERY SYSTEM
Manufacturer (Section D)
ABBOTT VASCULAR
26531 ynez rd.
temecula CA 92591 4628
Manufacturer (Section G)
ABBOTT VASCULAR, REG # 3005718570 (P099)
cashel road
clonmel tipperary
EI  
Manufacturer Contact
lindsey bell
26531 ynez rd.
temecula, CA 92591-4628
9519143996
MDR Report Key15240642
MDR Text Key304037245
Report Number2024168-2022-08825
Device Sequence Number1
Product Code NIQ
Combination Product (y/n)Y
Reporter Country CodeIN
PMA/PMN Number
P110019
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative,Distributor
Reporter Occupation Physician
Type of Report Initial
Report Date 08/17/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received08/17/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date11/05/2023
Device Catalogue Number1070300-38
Device Lot Number0110241
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/27/2022
Is the Reporter a Health Professional? Yes
Date Manufacturer Received07/28/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured11/06/2020
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-