It was reported that the customer had problems with foley statlocks (fol0101 - pr# (b)(4); fol0102 - pr# (b)(4)) not staying closed and the catheter tugging.The representative discussed with customer if they were using the correct statlock for their foley.Customer only knew it was a 2-way 18fr, but did not know the substrate or type.It was stated that the statlock only stayed closed for a few minutes and then popped open.The customer had also tried the leg strap option but that caused urethral and meatal trauma (pr# (b)(4)), stated they did not report the leg strap issue.It is unknown what medical intervention was provided.
|
It was reported that the customer had problems with foley statlocks (fol0101, fol0102) not staying closed and the catheter tugging.The representative discussed with customer if they were using the correct statlock for their foley.Customer only knew it was a 2-way 18fr, but did not know the substrate or type.It was stated that the statlock only stayed closed for a few minutes and then popped open.The customer had also tried the leg strap option but that caused urethral and meatal trauma, stated they did not report the leg strap issue.It is unknown what medical intervention was provided.
|
The reported issue could be inconclusive as no sample was returned for evaluation.A potential root cause for this failure could be ¿inappropriate snap fit¿.It was unknown whether the device had met specifications.The product was used for treatment purposes.It was unknown whether the product had caused the reported failure.The device was not returned for evaluation.The lot number is unknown; therefore, the device history record could not be reviewed.The instructions for use were found adequate and state the following: "indications for use: the statlock® device is a stabilization device for compatible catheters.Contraindications: known tape or adhesive allergies.Warnings and precautions: 1.Do not use the statlock® device where loss of adherence could occur, such as with a confused patient, diaphoretic or non-adherent skin, or when the access device is not monitored daily.2.Observe universal blood and body fluid precautions and infection control procedures, during application and removal of the statlock® device.3.Minimize catheter manipulation during application and removal of the statlock® device.4.Daily maintenance: a.The statlock® device should be assessed daily and changed when clinically indicated, at least every seven days.B.If pad becomes soiled, wash with soap/water, saline or hydrogen peroxide.Do not use alcohol or prepackaged bathing systems, which could lead to early lifting.C.If showering/bathing, cover with plastic wrap or waterproof dressing.D.Conduct skin assessment prior to application and repeat daily per facility protocol.E.Use clinical judgment on the removal of the statlock® stabilization device if the patient experiences any fluid shifts that may interfere with skin integrity.Application technique prep 1.Place foley catheter into retainer.Directional arrow should point towards catheter tip, and balloon inflation arm should be next to the clamp hinge.2.Close lid, being careful to avoid pinching the catheter.3.Identify securement site by laying the device retainer on the front of the thigh, leaving 1 inch of catheter slack between insertion site and the statlock® device retainer.4.After placing the statlock® stabilization device off to the side, cleanse and decrease the securement site with alcohol per hospital policy.Let skin dry.5.Apply skin protectant, in direction of hair growth, to area larger than securement site.Allow to dry completely (10-15 seconds).6.Using permanent marker, write initials and date of application on the statlock® device anchor pad.Note: always secure catheter into the statlock® device retainer before applying adhesive pad on skin.Place and peel 7.Align the statlock® stabilization device over securement site leaving 1 inch of catheter slack.Make sure leg is fully extended.8.While holding the retainer to keep the pad in place, peel away paper backing, one side at a time and place tension-free on skin.Removal technique: disengage 1.Open retainer by pressing release button with thumb, then lift to open.2.Remove foley catheter from the statlock® device.Dissolve: 3.Wipe the edge of the pad using at least 5-6 alcohol pads until a corner lifts.Then continue to stroke under surface of pad with alcohol to dissolve adhesive pad away from skin.Do not pull or force pad to remove." h11:section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.H3 other text : the device was not returned.
|