• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY MITEK LLC US 5 WAY FOOT PEDAL; FOOT-SWITCH, ELECTRICAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY MITEK LLC US 5 WAY FOOT PEDAL; FOOT-SWITCH, ELECTRICAL Back to Search Results
Model Number 283573
Device Problems Break (1069); Appropriate Term/Code Not Available (3191)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 07/21/2022
Event Type  malfunction  
Manufacturer Narrative
This report is being submitted in pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by depuy mitek or its employees that the report constitutes an admission that the device, depuy mitek, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.Udi: (b)(4).The complaint device was received at the service center and evaluated.During the service evaluation the following defects were identified: left pedal broken."blood stop" sticker to be replaced.Incorrect measured value due to cable default.Missing left magnet.Per service reports, this complaint can be confirmed.The cable , fischer cap, stickers , magnet were replaced to resolve the issues.After repair, the device was found to be working according to the specifications.As part of depuy mitek¿s quality process all devices are manufactured, inspected, and released to approved specifications.The improper maintenance was identified as the root cause for the device failure during the service evaluation.Additional complaint information monitoring for potential safety signals is conducted through complaint trending as part of post market surveillance.At this point in time, no corrective action is required, and no further action is warranted.Depuy mitek will continue to track any related complaints within this device family as a means of monitoring the extent with which this complaint is observed in the field.Manufacturing record evaluation is not required as the reported event is not associated with the manufacturing process and/or the potential cause of the defect cannot be associated to manufacturing.
 
Event Description
It was reported by a healthcare professional the the 5 way foot pedal had an unspecified malfunction.During in-house engineering evaluation, it was determined that the left pedal on the device was broken.There was no procedure nor patient involvement reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
5 WAY FOOT PEDAL
Type of Device
FOOT-SWITCH, ELECTRICAL
Manufacturer (Section D)
DEPUY MITEK LLC US
325 paramount drive
raynham MA 02767
Manufacturer (Section G)
DEPUY MITEK LLC US
325 paramount drive
raynham MA 02767
Manufacturer Contact
kate karberg
325 paramount drive
raynham, MA 02767
3035526892
MDR Report Key15241408
MDR Text Key305349103
Report Number1221934-2022-02524
Device Sequence Number1
Product Code HRX
UDI-Device Identifier10886705017010
UDI-Public10886705017010
Combination Product (y/n)N
Reporter Country CodeFR
PMA/PMN Number
K954465
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,User Facility
Reporter Occupation Other Health Care Professional
Type of Report Initial
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number283573
Device Catalogue Number283573
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/26/2022
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 08/05/2022
Initial Date FDA Received08/17/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured02/11/2020
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-