Device Problems Electrical /Electronic Property Problem (1198); Noise, Audible (3273)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Type  malfunction  
Manufacturer Narrative
The subject device has not been returned to olympus for evaluation.The investigation is ongoing, and a definitive root cause of the reported event cannot be determined at this time.If additional information becomes available, this report will be supplemented accordingly.
 
Event Description
The customer reported to olympus, at the end of a therapeutic (laparoscopic) procedure, when the stop button of the high flow insufflation unit was pressed, an error sound was generated, and the front panel blacked out.Subsequently, the device was turned off and on to resolve the issue.There was no harm or user injury reported due to the event.Additionally, the customer stated the issue has occurred previously.The previous incident is recorded under the medwatch report with patient identifier: (b)(6).
 
Manufacturer Narrative
This report is being supplemented to provide additional information based on the legal manufacturer's final investigation.It has been over 9 years since the subject device was manufactured.Based on the results of the legal manufacturer's investigation and based on a similar complaint with a returned product, it was determined that the display on the front panel was blacked out due to the failure of the main board.It was confirmed that when the stop button was pressed in an attempt to stop gas feeding of uhi-4, an error sound was generated, and the front panel blacked out.A review of the device history record found no deviations that could have caused or contributed to the reported issue.Olympus will continue to monitor field performance for this device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HIGH FLOW INSUFFLATION UNIT
Type of Device
HIGH FLOW INSUFFLATION UNIT
Manufacturer (Section D)
SHIRAKAWA OLYMPUS CO., LTD.
3-1 okamiyama
odakura, nishigo-mura,
nishishirakawa-gun, fukushima 961-8 061
JA  961-8061
Manufacturer Contact
masaharu hirose
3-1 okamiyama
odakura, nishigo-mura,
nishishirakawa-gun, fukushima 961-8-061
JA   961-8061
426422891
MDR Report Key15243562
MDR Text Key305368518
Report Number3002808148-2022-01112
Device Sequence Number1
Product Code HIF
UDI-Device Identifier04953170324147
UDI-Public04953170324147
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
K122180
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,Health Professional,User Facility,Distributor
Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date 10/27/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received08/17/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberUHI-4
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Date Manufacturer Received09/28/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured02/04/2013
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-