• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DORNOCH DUO FLUID CART WITH SMOKE EVACUATOR AND POWER IV; APPARATUS, SUCTION, WARD USE, PORTABLE, AC-POWERED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DORNOCH DUO FLUID CART WITH SMOKE EVACUATOR AND POWER IV; APPARATUS, SUCTION, WARD USE, PORTABLE, AC-POWERED Back to Search Results
Model Number N/A
Device Problems Thermal Decomposition of Device (1071); Melted (1385); Failure to Power Up (1476)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 07/07/2022
Event Type  malfunction  
Event Description
It was reported that this unit will not turn on and the fuses melted into the power inlet module.There were signs of fire, burning, and charring as the fuse was melted in the power inlet module.No adverse events were reported as a result of this malfunction.
 
Manufacturer Narrative
This event has been recorded by zimmer biomet under (b)(4).Investigation is completed.This is an initial final report submission.Review of the most recent repair record determined the unit would not turn on due to blown fuse in the power inlet module.The power inlet module was replaced and resolved the reported issue.Review of the device history record identified no deviations or anomalies during manufacturing related to the reported event.A definitive root cause cannot be determined.The event is confirmed.If any further information is found which would change or alter any conclusions or information, a supplemental report will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DUO FLUID CART WITH SMOKE EVACUATOR AND POWER IV
Type of Device
APPARATUS, SUCTION, WARD USE, PORTABLE, AC-POWERED
Manufacturer (Section D)
DORNOCH
200 northwest parkway
riverside MO 64150
Manufacturer (Section G)
DORNOCH
200 northwest parkway
riverside MO 64150
Manufacturer Contact
jennifer rapsavage
56 e. bell drive
warsaw, IN 46582
5745260384
MDR Report Key15244559
MDR Text Key298106820
Report Number0001954182-2022-00079
Device Sequence Number1
Product Code JCX
UDI-Device Identifier00889024501638
UDI-Public(01)00889024501638
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K190789
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type User Facility,Company Representative
Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date 08/17/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberN/A
Device Catalogue Number00514010400
Device Lot Number44604
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 08/01/2022
Initial Date FDA Received08/17/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured12/22/2020
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Removal/Correction NumberN/A
Patient Sequence Number1
-
-