• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MAQUET CRITICAL CARE AB SUPPORT ARM 177; SUPPORT, ARM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MAQUET CRITICAL CARE AB SUPPORT ARM 177; SUPPORT, ARM Back to Search Results
Model Number SUPPORT ARM 177
Device Problem Fracture (1260)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 07/28/2022
Event Type  malfunction  
Manufacturer Narrative
No repair was requested and no parts were returned.The support arm is a casting and it most probably developed a crack at an earlier occasion either by overloading or impact and this led to the reported breaking.The support arm has been successfully tested for mechanical strength and is designed according to standard.Previous investigations of similar faults led to a redesign and change of the manufacturing process in order to obtain a higher mechanical strength of the support arm.The support arm associated with this complaint had been manufactured before this change was implemented in production.
 
Event Description
It was reported that the support arm for patient tubes broke.There was no patient harm.Manufacturer's ref #: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SUPPORT ARM 177
Type of Device
SUPPORT, ARM
Manufacturer (Section G)
MAQUET CRITICAL CARE AB
roentgenvagen 2
solna
Manufacturer Contact
sara soderholm roman
roentgenvagen 2
solna 
MDR Report Key15247527
MDR Text Key298163460
Report Number8010042-2022-01423
Device Sequence Number1
Product Code IOY
Combination Product (y/n)N
Reporter Country CodePO
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Company Representative
Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date 08/18/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberSUPPORT ARM 177
Device Catalogue Number6481720
Was Device Available for Evaluation? No
Initial Date Manufacturer Received 07/28/2022
Initial Date FDA Received08/18/2022
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-