• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: TERUMO BCT TRIMA ACCEL; TRIMA ACCEL AUTOMATED BLOOD COLLECTION SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

TERUMO BCT TRIMA ACCEL; TRIMA ACCEL AUTOMATED BLOOD COLLECTION SYSTEM Back to Search Results
Model Number 81012
Device Problems Mechanical Problem (1384); Device Slipped (1584)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 07/27/2022
Event Type  malfunction  
Manufacturer Narrative
Investigation: the iv pole safety collar was present on the device prior to the reported incident.A service engineer adjusted the front door latch and the iv pole lock release button.The device serial number history report indicates no further related issues have been reported for this device.One year of service history was reviewed for this device with no issues related to the reported condition identified.Root cause: a root cause assessment was performed for this complaint.The root cause for the door lock and iv pole issues was determined to be the need for an alignment/adjustment.
 
Event Description
During a service call, the terumo bct technician found that the iv pole drops when the the side panel was pressed.There was not a patient or operator involved at the time the iv pole fell, therefore patient information is not reasonably known.This product is not available within the us, but this report is being filed due to an alleged failure that could occur on a similarly marketed device platform cleared for use by the fda.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TRIMA ACCEL
Type of Device
TRIMA ACCEL AUTOMATED BLOOD COLLECTION SYSTEM
Manufacturer (Section D)
TERUMO BCT
lakewood CO 80215
Manufacturer (Section G)
TERUMO BCT
10810 w. collins ave
lakewood CO 80215
Manufacturer Contact
scot hilden
10810 w. collins ave
lakewood, CO 80215
MDR Report Key15251385
MDR Text Key298167290
Report Number1722028-2022-00273
Device Sequence Number1
Product Code GKT
Combination Product (y/n)N
Reporter Country CodeJA
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 08/18/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received08/18/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number81012
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Date Manufacturer Received07/27/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured07/11/2018
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
Patient Outcome(s) Other;
-
-