W. L. GORE & ASSOCIATES, INC. GORE® TAG® CONFORMABLE THORACIC STENT GRAFT WITH ACTIVE CONTROL SYSTEM; SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT Back to Search Results
Catalog Number TGM343420J
Device Problem No Apparent Adverse Event (3189)
Patient Problem Aneurysm (1708)
Event Date 07/21/2022
Event Type  Injury  
Manufacturer Narrative
(b)(4).W.L.Gore & associates, inc.(gore) is submitting this report to comply with 21 c.F.R.Part 803, the medical device reporting regulation.This report is based upon information obtained by gore, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Blank fields present on this report include required fields and fields determined to be not applicable.Blank required fields indicate that the information was not provided, was deemed unavailable or was not applicable.This report does not constitute an admission or a conclusion by fda, gore, or its associates that the device, gore or its associates caused or contributed to the event described in the report.In particular, this report does not constitute a legal admission by anyone that the product described in this report has any defects or has malfunctioned, as defined from a legal standpoint.These words are included in the report and are fixed items for selection created by the fda, to categorize the type of event solely for the purpose of reporting pursuant to part 803.This statement should be included with any information or report disclosed to the public under the freedom of information act.
 
Event Description
On (b)(6) 2021, a patient underwent endovascular treatment of a thoracic aortic disease using gore® tag® conformable thoracic stent graft with active control system (ctagac), tgm343420j.The patient had already undergone a total arch replacement and a frozen elephant trunk.The ctagac was implanted to extend distally from the elephant trunk.On unknown date (remains unknown), a pseudoaneurysm was observed at the distal end of the ctagac.On (b)(6) 2022, a reintervention was performed to treat the pseudoaneurysm.A gore® tag® conformable thoracic stent graft with active control system (ctagac), tgmr313120j, was implanted distally.A proximal type i endoleak of tgmr313120j was observed.To treat it, additionally gore® tag® conformable thoracic stent graft with active control system (ctagac), tgm343415j, was implanted proximally.A minor proximal type i endoleak was remained at an outer curvature.The physician decided to monitor it.The patient tolerated the procedure.The physician stated that regarding the pseudoaneurysm, the ctagac device might be oversized.Reportedly, tmg343420j might be used to treat a distal stent graft-induced new entry(sine) at the elephant trunk.But it was not confirmed.A vessel diameter at a location of the distal end of tgm343420j was originally 29 mm.Medical history: total arch replacement, frozen elephant trunk.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GORE® TAG® CONFORMABLE THORACIC STENT GRAFT WITH ACTIVE CONTROL SYSTEM
Type of Device
SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT
Manufacturer (Section D)
W. L. GORE & ASSOCIATES, INC.
1505 n. fourth street
flagstaff AZ 86004
Manufacturer (Section G)
MEDICAL WOODY SPRINGS B/P
3450 w. kiltie lane
flagstaff AZ 86005
Manufacturer Contact
samir kulovic
1505 n. fourth street
flagstaff, AZ 86004
9285263030
MDR Report Key15255475
MDR Text Key298193768
Report Number2017233-2022-03239
Device Sequence Number1
Product Code MIH
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
P040043
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Company Representative
Reporter Occupation Physician
Type of Report Initial
Report Date 08/19/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Expiration Date11/23/2023
Device Catalogue NumberTGM343420J
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 07/21/2022
Initial Date FDA Received08/18/2022
Was Device Evaluated by Manufacturer? No
Date Device Manufactured11/23/2020
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient Age76 YR
Patient SexMale
-
-