Model Number N/A |
Device Problems
Corroded (1131); Naturally Worn (2988)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 04/25/2022 |
Event Type
malfunction
|
Manufacturer Narrative
|
(b)(4).Report source - foreign: france.The results of the investigation are as follows: review of the most recent repair record determined the screws were worn and motor speed unstable and motor and screws were replaced and resolved the reported issue.A definitive root cause cannot be determined.The event is confirmed.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
|
|
Event Description
|
It was reported during maintenance that the following failure descriptions were noted: worn screws need to be replaced; corroded motor need to be replaced.Upon investigation it was noted that there was inconsistent/unstable speed in the motor.There was no patient involvement.Due diligence is complete as multiple attempts were made; however, no further information is available.As no additional information or product is available, we are unable to provide further information.
|
|