MAUDE Adverse Event Report: DORNOCH DUO FLUID CART WITH SMOKE EVACUATOR AND POWER IV; APPARATUS, SUCTION, WARD USE, PORTABLE, AC-POWERED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DORNOCH DUO FLUID CART WITH SMOKE EVACUATOR AND POWER IV; APPARATUS, SUCTION, WARD USE, PORTABLE, AC-POWERED Back to Search Results
Model Number N/A
Device Problem Smoking (1585)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 07/20/2022
Event Type  malfunction  
Manufacturer Narrative
(b)(4).The results of the investigation are as follows: review of the most recent repair record determined the fuse was bad and had damaged the power inlet module and the power inlet module was replaced and resolved the reported issue.A definitive root cause cannot be determined.The event is confirmed.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Event Description
It was reported the power inlet module was burnt and it burnt the power cord.The event timing was not during surgery.No patient involvement or harm.No further event information available at the time of this report.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DUO FLUID CART WITH SMOKE EVACUATOR AND POWER IV
Type of Device
APPARATUS, SUCTION, WARD USE, PORTABLE, AC-POWERED
Manufacturer (Section D)
DORNOCH
200 northwest parkway
riverside MO 64150
Manufacturer (Section G)
DORNOCH
200 northwest parkway
riverside MO 64150
Manufacturer Contact
jennifer rapsavage
56 e. bell drive
warsaw, IN 46582
5745260384
MDR Report Key15257813
MDR Text Key303061640
Report Number0001954182-2022-00080
Device Sequence Number1
Product Code JCX
UDI-Device Identifier00889024501638
UDI-Public(01)00889024501638
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K190789
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type User Facility,Company Representative
Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date 08/19/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received08/19/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberN/A
Device Catalogue Number00514010400
Device Lot Number45485
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Date Manufacturer Received08/10/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured05/17/2021
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Removal/Correction NumberN/A