This mdr report product is for an ous event identified as same/similar during a retrospective review as a result of abbott¿s investigation.The reported event of body fluid inside the lv lead bore was confirmed in the lab.However, the existence of body fluid inside the lead bores is considered normal, given the device had been implanted for almost 9 years.The set screws were fully secured and functioned properly during bench test.Interrogation of the device revealed the device was above elective replacement indicator (eri) when received.Telemetry, pacing, sensing, impedance, hv output, hv shock, and patient notifier were tested and no anomalies were found.
|