Model Number IPN000302
Device Problems Device Alarm System (1012); Mechanical Problem (1384); Connection Problem (2900)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 02/09/2022
Event Type  malfunction  
Manufacturer Narrative
Qn#(b)(4).
 
Event Description
It was reported that "during use, the system error 3 alarm, the drive board is poorly connected, and the engineer replaces the drive board".Additional information received stated that the pump was changed.No patient harm, injury, or consequence reported.Patient condition reported as "fine".
 
Manufacturer Narrative
(b)(4).The reported complaint of system error 3 alarm is not able to be confirmed.The product was not returned for investigation.According to the additional information received, the m-force motor driver was replaced.Based on a review of the device history record (dhr), the product met specification upon release.The root cause of the complaint is undetermined.No further action required at this time.The reported complaint will be monitored for any developing trends.
 
Event Description
It was reported that "during use, the system error 3 alarm, the drive board is poorly connected, and the engineer replaces the drive board".Additional information received stated that the pump was changed.No patient harm, injury, or consequence reported.Patient condition reported as "fine".
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AUTOCAT 2
Type of Device
SYSTEM, BALLOON, INTRA-AORTIC
Manufacturer (Section D)
ARROW INTERNATIONAL LLC
morrisville NC
Manufacturer (Section G)
ARROW INTERNATIONAL INC.
16 elizabeth drive
chelmsford MA 01824
Manufacturer Contact
marling heaton
3015 carrington mill blvd
morrisville 27560
MDR Report Key15266894
MDR Text Key305247885
Report Number3010532612-2022-00317
Device Sequence Number1
Product Code DSP
UDI-Device Identifier30801902051722
UDI-Public30801902051722
Combination Product (y/n)N
Reporter Country CodeCH
PMA/PMN Number
K060309
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,User Facility,Company Representative
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 07/26/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Model NumberIPN000302
Device Catalogue NumberIAP-0400
Device Lot NumberN/A
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 07/26/2022
Initial Date FDA Received08/22/2022
Supplement Dates Manufacturer Received09/14/2022
Supplement Dates FDA Received09/16/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
Treatment
NONE REPORTED; NONE REPORTED
-
-