Model Number IPN000302 |
Device Problems
Device Alarm System (1012); Mechanical Problem (1384); Connection Problem (2900)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 02/09/2022 |
Event Type
malfunction
|
Manufacturer Narrative
|
Qn#(b)(4).
|
|
Event Description
|
It was reported that "during use, the system error 3 alarm, the drive board is poorly connected, and the engineer replaces the drive board".Additional information received stated that the pump was changed.No patient harm, injury, or consequence reported.Patient condition reported as "fine".
|
|
Manufacturer Narrative
|
(b)(4).The reported complaint of system error 3 alarm is not able to be confirmed.The product was not returned for investigation.According to the additional information received, the m-force motor driver was replaced.Based on a review of the device history record (dhr), the product met specification upon release.The root cause of the complaint is undetermined.No further action required at this time.The reported complaint will be monitored for any developing trends.
|
|
Event Description
|
It was reported that "during use, the system error 3 alarm, the drive board is poorly connected, and the engineer replaces the drive board".Additional information received stated that the pump was changed.No patient harm, injury, or consequence reported.Patient condition reported as "fine".
|
|