• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: NEUROPACE, INC. NEUROPACE RNS SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

NEUROPACE, INC. NEUROPACE RNS SYSTEM Back to Search Results
Model Number DL-330-10-K
Device Problems Signal Artifact/Noise (1036); Impedance Problem (2950)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 08/01/2022
Event Type  malfunction  
Manufacturer Narrative
(b)(4).The explanted product was not returned to neuropace for analysis.
 
Event Description
On (b)(6) 2022, during an ecog review, the patient's lead showed evidence of a lead fracture.On (b)(6) 2022 the lead was replaced.There is no information suggesting the cause of the break.The lead was secured with dog bone with lead protection and had been implanted for 1.4 years prior to explant.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NEUROPACE RNS SYSTEM
Type of Device
NEUROPACE RNS SYSTEM
Manufacturer (Section D)
NEUROPACE, INC.
455 n. bernardo ave.
mountain view CA 94043
Manufacturer (Section G)
NEUROPACE, INC.
455 n. bernardo ave.
mountain view CA 94043
Manufacturer Contact
ramona gonis
455 n. bernardo ave.
mountain view, CA 94043
6502382788
MDR Report Key15271658
MDR Text Key305538141
Report Number3004426659-2022-00026
Device Sequence Number1
Product Code PFN
UDI-Device Identifier00855547005014
UDI-Public010085554700501417220823
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P100026
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date 08/22/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received08/22/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Model NumberDL-330-10-K
Device Catalogue Number1007607
Device Lot Number28803-1-1-1
Date Manufacturer Received08/01/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Other;
Patient Age39 YR
Patient SexMale
-
-