• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SYNCARDIA SYSTEMS, LLC SYNCARDIA FREEDOM DRIVER; EXTERNAL PNEUMATIC DRIVER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SYNCARDIA SYSTEMS, LLC SYNCARDIA FREEDOM DRIVER; EXTERNAL PNEUMATIC DRIVER Back to Search Results
Catalog Number 595000-001
Device Problems Fracture (1260); No Apparent Adverse Event (3189); Audible Prompt/Feedback Problem (4020)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 08/10/2022
Event Type  malfunction  
Manufacturer Narrative
The freedom driver will be returned to syncardia for evaluation.The results will be provided in a follow-up mdr.
 
Event Description
The customer, a syncardia authorized distributor, reported that the freedom driver exhibited a fault alarm while supporting a patient.There was no reported adverse patient impact.The patient was in the hospital in bed because of raised body temperature, with good blood pressure, and normal fill volumes and cardiac output values.The customer also reported that the patient was subsequently switched to a backup driver.
 
Manufacturer Narrative
Visual inspection found no evidence of secondary motor engagement.Visual inspection found a crack on the driver display cover.Damage is cosmetic and not related to the customer reported issue.No other damage to internal or external components was found.The driver underwent extended observation testing in an attempt to duplicate the reported fault alarm; no issues or alarms were observed during the observation test.The customer reported fault alarm could not be confirmed or replicated.No evidence of a device malfunction was found during the investigation.This issue will be monitored and trended as part of the customer complaint process.Syncardia has completed its evaluation and is closing this file.If new or additional information is received in the future, syncardia will file a follow-up mdr.(b)(4) follow-up report 1.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SYNCARDIA FREEDOM DRIVER
Type of Device
EXTERNAL PNEUMATIC DRIVER
Manufacturer (Section D)
SYNCARDIA SYSTEMS, LLC
1992 e. silverlake road
tucson AZ 85713
Manufacturer (Section G)
SYNCARDIA SYSTEMS, LLC
1992 e. silverlake road
tucson AZ 85713
Manufacturer Contact
aaron meier
1992 e. silverlake road
tucson, AZ 85713
5205451234
MDR Report Key15272203
MDR Text Key298581846
Report Number3003761017-2022-00089
Device Sequence Number1
Product Code LOZ
Combination Product (y/n)N
PMA/PMN Number
P030011
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Distributor
Reporter Occupation Administrator/Supervisor
Type of Report Initial,Followup
Report Date 12/22/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Catalogue Number595000-001
Was Device Available for Evaluation? Device Returned to Manufacturer
Initial Date Manufacturer Received 08/10/2022
Initial Date FDA Received08/22/2022
Supplement Dates Manufacturer Received08/10/2022
Supplement Dates FDA Received12/28/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured03/20/2019
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
Patient Age44 YR
Patient SexMale
-
-