• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ST PAUL HOTLINE 3 BLOOD AND FLUID WARMER; WARMER, THERMAL, INFUSION FLUID

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ST PAUL HOTLINE 3 BLOOD AND FLUID WARMER; WARMER, THERMAL, INFUSION FLUID Back to Search Results
Catalog Number HL-390
Device Problem Electrical /Electronic Property Problem (1198)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Type  malfunction  
Manufacturer Narrative
A product sample was received for evaluation.Visual inspection and functional testing were performed.The enclosure, front cover, line cord and pole clamp were damaged.During functional testing, the technician was able to duplicate the reported issue of failed ground bond/ electrical test.The technician found out that one of the ground wires was loose and the nut securing the loose ground wire was tightened to resolve the reported issue.All the damaged parts were replaced.The technician performed the preventive maintenance (pm).The device passed all the functional testing.The root cause of failed ground bond/ electrical test was determined to be the loose ground wire.Manufacturing device history record review was not performed because the results of the complaint investigation do not indicate a problem with the manufacture of the device.(udi number) is unknown.No product information has been provided to date.This remediation mdr was generated under protocol (b)(4), as a result of warning letter cms# (b)(4).
 
Event Description
It was reported that the device failed the ground bond/ electrical test.No patient injury was reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HOTLINE 3 BLOOD AND FLUID WARMER
Type of Device
WARMER, THERMAL, INFUSION FLUID
Manufacturer (Section D)
ST PAUL
1265 grey fox rd.
st. paul MN 55112
Manufacturer (Section G)
ST PAUL
1265 grey fox rd.
st. paul MN 55112
Manufacturer Contact
jim vegel
6000 nathan lane north
minneapolis, MN 55442
MDR Report Key15276903
MDR Text Key304987121
Report Number3012307300-2022-15853
Device Sequence Number1
Product Code LGZ
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K911383
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type User Facility
Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date 08/23/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberHL-390
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/04/2019
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 04/12/2022
Initial Date FDA Received08/23/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured03/18/2014
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-