• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: TERUMO PHILIPPINES CORPORATION MCKESSON PREVENT ® HT SAFETY HYPODERMIC NEEDLE; SYRINGE, ANTISTICK

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

TERUMO PHILIPPINES CORPORATION MCKESSON PREVENT ® HT SAFETY HYPODERMIC NEEDLE; SYRINGE, ANTISTICK Back to Search Results
Model Number N/A
Device Problem Activation Problem (4042)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 07/13/2022
Event Type  malfunction  
Manufacturer Narrative
Lot number: requested, unknown.Expiration date: unknown due to unknown lot number.Udi: unknown due to unknown lot number.Implanted date: device was not implanted.Explanted date: device was not explanted.Device manufacture date: unknown due to unknown lot number the actual sample was not available for our evaluation.The root cause of the complaint could not be identified to be related to our product or production process.Lot history file and retention samples were unable to confirm since the lot of this complaint is unknown.Instead, simulation through manual sheath activation was conducted on the current lot samples.The needle is fully engaged under the lock (sheath tooth) and no irregularity during and after activation was encountered that may lead to the complaint.We have a series of visual in-process inspections to detect an abnormality on the sheath that may lead to a problem during sheath activation.The molding condition of the components critical to the safety activation of the product is routinely checked to assure that no defects will be encountered that will lead to sheath activation problems and needle sticks.Similarly, the assembly status of the safety needle such as sheath collar fitting and damaged parts that will affect product function is being confirmed.Before shipment, qc conducts outgoing visual, sensory, and functional inspections to assure lots are of good quality.Only samples passed results are allowed to be shipped.Therefore, we advise to follow the instructions for use (ifu) for the proper usage of the sg2 needle indicated on the unit box in which warnings to prevent needle stick, cautions, and precautions are also included.(b)(4).
 
Event Description
The user facility nurse reported that there was a safety issue with the terumo needle.The cover guard would not stay closed, or latch closed which increased the risk of a needle stick occurring.Additional information was received on 03 aug 2022: it was reported that the issue with the needle was that they couldn't get the safety sheath to stay closed after attempting to activate.Activation was against a hard surface.It was difficult during activation, they heard a click when it was activated, but when the needle was disposed in container, the needle was totally exposed, and had not stayed in the sheath guard.A needle stick did not occur and there were no staff injuries.No patients were affected in any way, as they received their injections as intended.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MCKESSON PREVENT ® HT SAFETY HYPODERMIC NEEDLE
Type of Device
SYRINGE, ANTISTICK
Manufacturer (Section D)
TERUMO PHILIPPINES CORPORATION
124 east main avenue
laguna technopark
binan,
RP 
Manufacturer (Section G)
TERUMO PHILIPPINES CORPORATION
124 east main avenue
laguna technopark
binan, laguna,,
RP  
Manufacturer Contact
gina digioia
265 davidson ave
suite 320
somerset, NJ 08873
6402040886
MDR Report Key15278035
MDR Text Key305498577
Report Number3003902955-2022-00030
Device Sequence Number1
Product Code MEG
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K051865
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type User Facility,Company Representative
Reporter Occupation Nurse
Type of Report Initial
Report Date 08/23/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received08/23/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberN/A
Device Catalogue Number102-N231S
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received07/29/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-