• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: TELEFLEX MEDICAL HEMOLOK ML CLIPS 3/CART 42/BOX; CLIP, IMPLANTABLE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

TELEFLEX MEDICAL HEMOLOK ML CLIPS 3/CART 42/BOX; CLIP, IMPLANTABLE Back to Search Results
Catalog Number 544233
Device Problem Break (1069)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 08/15/2022
Event Type  malfunction  
Event Description
Hemolok was broken during use on surgery.
 
Manufacturer Narrative
Qn#(b)(4).The device history review for the product hemolok ml clips 3/cart 42/box lot# 73f2100268 investigation did not show issues related to the complaint.
 
Manufacturer Narrative
Qn#(b)(4).The customer returned one cartridge with one broken clip from one unit of 544233 hemolok ml clips 6/cart 42/box for investigation.The returned components were visually examined with and without magnification.Visual examination revealed that the loose clip was broken in half at the hinge.Small scrapes were observed on the returned cartridge.No other defects or anomalies were observed.Evidence of use in the form of biological material was observed on the returned clip.The clip breaking in half at the hinge during ligation was determined to be the result of insert mismatch at the hinge area of the clip.The root cause is due to inadequate specifications for insert mismatch in the highest stress area of the hinge, which is a design related issue.A capa has been previously opened to further investigate issues related to clip breakages.The ifu for this product, l02425, was reviewed as a part of this complaint investigation.The ifu states, "always check the alignment of the applier jaws before use.When closed, jaw tips should be directly aligned and not offset.Alignment of the jaw is critical for safe application of the clip.If this is not done, patient injury may occur.Proper maintenance, care and cleaning are necessary to ensure proper functionality." in summary, the reported complaint of "broken/detached parts - clip - hinge" was confirmed based upon the sample received.A broken clip was returned, and it was broken in half at the hinge.The clip breaking at the hinge during ligation was determined to be the result of insert mismatch at the hinge area on the pierce leg side of the hinge.The root cause is due to inadequate specifications for insert mismatch in the highest stress area of the hinge, which is a design related issue.A capa has been previously opened to further investigate issues related to clip breakages.Teleflex will continue to monitor and trend on complaints of this nature.
 
Event Description
Hemolok was broken during use on surgery.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HEMOLOK ML CLIPS 3/CART 42/BOX
Type of Device
CLIP, IMPLANTABLE
Manufacturer (Section D)
TELEFLEX MEDICAL
morrisville NC
Manufacturer (Section G)
TELEFLEX MEDICAL
rancho el descanso
tecate 21478
MX   21478
Manufacturer Contact
jasmine brown
3015 carrington mill blvd
morrisville, NC 27560
9193614124
MDR Report Key15279370
MDR Text Key304982095
Report Number3003898360-2022-00404
Device Sequence Number1
Product Code FZP
Combination Product (y/n)N
Reporter Country CodeTW
PMA/PMN Number
K133202
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 08/15/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received08/23/2022
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number544233
Device Lot Number73F2100268
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Date Manufacturer Received09/23/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured06/08/2021
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage A
Patient Sequence Number1
-
-