Model Number V60
Device Problems Leak/Splash (1354); Volume Accuracy Problem (1675)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 08/03/2022
Event Type  malfunction  
Event Description
It was reported that the ventilator during set up had no tidal volume.There was no patient involvement.The customer troubleshot with technical support and was advised to run unit on test circuit at set parameters and try to duplicate event, tidal volume (vt) display should populate within two breaths with no alarms activated.The customer was also advised to perform performance verification test (pvt).Further information has been requested.If additional information becomes available at a later date, a supplemental report will be submitted.
 
Manufacturer Narrative
H10: multiple good faith efforts were made to retrieve device evaluation, repair, and operational status on 08/09/2022, 08/23/2022 and 09/21/2022, however, yielded no response from the customer.It is unknown if any parts or repairs have been conducted.The complaint will be processed for closure.If additional information is received at a later date, the complaint will be reopened, and a supplemental report will be submitted.
 
Manufacturer Narrative
Upon a follow up with technical support the customer reported that the unit had an error code indicating low leak-co2 rebreathing risk.The customer was advised to replace the flow sensor.The customer had requested an onsite service.Unit is pending repair.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RESPIRONICS
Type of Device
VENTILATOR, CONTINUOUS, MINIMAL VENTILATORY SUPPORT, FACILITY USE
Manufacturer (Section D)
RESPIRONICS CALIFORNIA, LLC
2271 cosmos court
carlsbad CA 92011
Manufacturer (Section G)
RESPIRONICS CALIFORNIA, LLC
2271 cosmos court
carlsbad CA 92011
Manufacturer Contact
kimberly shelly
2271 cosmos court
carlsbad, CA 92011
7609187300
MDR Report Key15282670
MDR Text Key301130105
Report Number2031642-2022-02220
Device Sequence Number1
Product Code MNT
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K102985
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility
Reporter Occupation Biomedical Engineer
Type of Report Initial,Followup,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberV60
Device Catalogue Number1053617
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 08/03/2022
Initial Date FDA Received08/24/2022
Supplement Dates Manufacturer Received08/12/2022
09/29/2022
Supplement Dates FDA Received09/09/2022
09/30/2022
Date Device Manufactured02/20/2014
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-