• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: STRYKER ENDOSCOPY-SAN JOSE PKG, CROSSFIRE 2 CONSOLE; ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STRYKER ENDOSCOPY-SAN JOSE PKG, CROSSFIRE 2 CONSOLE; ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES Back to Search Results
Model Number 0475100000
Device Problems Sparking (2595); Temperature Problem (3022)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 08/09/2022
Event Type  malfunction  
Manufacturer Narrative
Additional information will be provided once the investigation has been completed.The device manufacturer date is not known at this time.However, should it become available it will be provided in future reports.
 
Event Description
It was reported that there was a thermal event.Please note this was not during a procedure.
 
Event Description
It was reported that there was a thermal event.Please note this was not during a procedure.
 
Manufacturer Narrative
This complaint investigation was closed based on the device not received, therefore the reported failure mode was not confirmed.In the event that the device is received, the complaint will be reopened and the investigation will be updated with new results.Alleged failure: sparked.Probable root cause: console fuses fail to break circuit due to excessive mains current leading to fire isolation power supply failure.Use error: console is sterilized or disinfected.Use error: console cleaned with incompatible products.The reported failure mode will be monitored for future reoccurrence.Manufacture date is not known.H3 other text : 81.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PKG, CROSSFIRE 2 CONSOLE
Type of Device
ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES
Manufacturer (Section D)
STRYKER ENDOSCOPY-SAN JOSE
5900 optical court
san jose CA 95138
Manufacturer (Section G)
STRYKER ENDOSCOPY-SAN JOSE
5900 optical court
san jose CA 95138
Manufacturer Contact
andrea zenere
5900 optical court
san jose, CA 95138
4087542000
MDR Report Key15293783
MDR Text Key300909443
Report Number0002936485-2022-00483
Device Sequence Number1
Product Code GEI
UDI-Device Identifier07613327058109
UDI-Public07613327058109
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K071859
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Company Representative
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 10/22/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received08/25/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number0475100000
Device Catalogue Number0475100000
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Date Manufacturer Received08/09/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-