Catalog Number 03666794001 |
Device Problem
Image Display Error/Artifact (1304)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 07/29/2022 |
Event Type
malfunction
|
Event Description
|
It was reported that there was an issue with the display of coaguchek xs meter serial number (b)(4).Segments were missing from the results field of the display.A display check was performed and in the results field, it displayed "000" on the bottom part of the screen, rather than a complete "888".In place of the top part of the "888" are 3 horizonal lines.More than one segment of the display was missing.The meter has not been dropped or broken.
|
|
Manufacturer Narrative
|
The customer's meter was requested for investigation and a replacement was sent to the customer.Occupation - the occupation is patient/consumer.
|
|
Manufacturer Narrative
|
The meter was returned for investigation.When examining the display, segments are shown with a weak contrast.The battery contacts in the battery compartment and circuit board were contaminated by a liquid (leaked battery) which has penetrated/corroded the solder contacts.The contamination would be caused by a mishandling of the device.
|
|