Other, other text: h6: updated.H10: device evaluation: the device was returned for investigation.Visual inspection and functional test were performed.The device was found to be missing the tamper seal.The event log showed no evidence of the reported problem.Upon function testing, the reported problem could not be duplicated as the device operated as intended.The root cause of the reported problem is unknown.The software was reinstalled as a preventative measure.A manufacturing device history record review was not performed because the results of the complaint investigation do not indicate a problem with the manufacture or last repair of the device.No manufacturing or service issues were identified as causes of the customer's reported problem during the review of service and repair records.Therefore, no manufacturing or service records review is needed.
|