Model Number 21-7302-24
Device Problem Failure to Deliver (2338)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Type  malfunction  
Event Description
It was reported that the infusion did not achieve patient when there was medicine in the cassette.Changing the cassette fixed the problem.No patient injury was reported.
 
Manufacturer Narrative
A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 if additional reportable information becomes available.Operator of device is unknown.No product information has been provided to date.
 
Manufacturer Narrative
Other text: h6: health impact and evaluation codes: updated.H10: a device history record (dhr) review was conducted which indicated all inspections were completed and no issues were noted during manufacture.No product sample nor photos were received; therefore, visual and functional testing could not be performed.The reported issue could not be confirmed.If the product is returned, the manufacturer will reopen this complaint for further investigation.This remediation mdr was generated under protocol (b)(4), as a result of warning letter cms# (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CADD ADMINISTRATION SET WITH FLOW STOP, MEDICATION
Type of Device
SET, ADMINISTRATION, INTRAVASCULAR
Manufacturer (Section D)
ST PAUL
1265 grey fox rd.
st. paul MN 55112
Manufacturer (Section G)
NULL
1265 grey fox rd.
st. paul MN 55112
Manufacturer Contact
jim vegel
6000 nathan lane north
minneapolis, MN 55442
MDR Report Key15300314
MDR Text Key305364606
Report Number3012307300-2022-16357
Device Sequence Number1
Product Code FPA
UDI-Device Identifier10610586027239
UDI-Public10610586027239
Combination Product (y/n)N
Reporter Country CodeTU
PMA/PMN Number
K040636
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Distributor
Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date 05/12/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received08/26/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number21-7302-24
Device Catalogue Number21-7302-24
Device Lot Number4219997
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Date Manufacturer Received04/10/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured01/11/2022
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-