MAUDE Adverse Event Report: AIZU OLYMPUS CO., LTD. HIGH FLOW INSUFFLATION UNIT; LAPAROSCOPIC INSUFFLATOR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

AIZU OLYMPUS CO., LTD. HIGH FLOW INSUFFLATION UNIT; LAPAROSCOPIC INSUFFLATOR Back to Search Results
Model Number UHI-4
Device Problems Intermittent Continuity (1121); No Device Output (1435)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Type  malfunction  
Event Description
During an asset return inspection with no complaint from the customer, an error code e03 was observed on the device.There was no patient involvement on this reported event.No user injury reported.
 
Manufacturer Narrative
Device evaluation and inspection, service repair noted the following findings: error e03 observed intermittently likely due to faulty main board (printed circuit board failure).Additionally, housing inspection found scratches on top cover.Investigation activities have been opened to manage the actions related to this issue and any required mdr reporting.¿ investigation is ongoing.This report will be supplemented accordingly following investigation.
 
Manufacturer Narrative
This report is being supplemented to provide additional information based on the legal manufacturer's final investigation.The investigation confirmed the reported e03 error code which indicates an abnormal pressure sensor.A review of the device history record (dhr) found no deviations that could have caused or contributed this issue.The dhr confirmed that the subject device was shipped in accordance with the specifications.A definitive root cause for this issue was not established.However, it is probable that the issue occurred because of the main internal circuit board was faulty.The root cause of the fault was unknown.Olympus will continue to monitor field performance for this device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HIGH FLOW INSUFFLATION UNIT
Type of Device
LAPAROSCOPIC INSUFFLATOR
Manufacturer (Section D)
AIZU OLYMPUS CO., LTD.
3-1 okamiyama
odakura, nishigo-mura
nishishirakawa-gun, fukushima 961-8 061
JA  961-8061
Manufacturer Contact
masaharu hirose
3-1 okamiyama
odakura, nishigo-mura,
nishishirakawa-gun, fukushima 961-8-061
JA   961-8061
426422891
MDR Report Key15303703
MDR Text Key305507489
Report Number3002808148-2022-01439
Device Sequence Number1
Product Code HIF
UDI-Device Identifier04953170324147
UDI-Public04953170324147
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K122180
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Company Representative
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 10/14/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberUHI-4
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/15/2022
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 07/15/2022
Initial Date FDA Received08/26/2022
Supplement Dates Manufacturer Received09/16/2022
Supplement Dates FDA Received10/14/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured05/27/2012
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-