• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ETHICON ENDO-SURGERY, LLC. LIGAMAX-5MM ENDO CLIP APPLIER; CLIP, IMPLANTABLE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ETHICON ENDO-SURGERY, LLC. LIGAMAX-5MM ENDO CLIP APPLIER; CLIP, IMPLANTABLE Back to Search Results
Model Number EL5ML
Device Problem Insufficient Information (3190)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 07/13/2022
Event Type  malfunction  
Manufacturer Narrative
(b)(4).Date sent: (b)(4).Batch # x94r53.Additional information was requested and the following was received: please clarify how the ¿clips could not be fed into the jaw properly¿.Did device not feed clips? no further information will be available.Did device feed clips sideways? no further information will be available.Did device not fire clips (jammed)? no further information will be available.Did device fire malformed clips? no further information will be available.Did device fire scissored clips? no further information will be available.Did device drop or eject clips? if other, please specify.No further information will be available.Investigation summary: the product was returned to ethicon for evaluation.Visual inspection and functional testing were conducted on the returned device.A manufacturing record evaluation was performed for the finished device batch number, and no non-conformances were identified.Visual analysis of the returned sample determined that the el5ml device was received with no damage in the external components.In an attempt to replicate the reported incident, the device was tested for functionality.Upon testing, the device was cycled and it fed and form 1 conforming clip and in the next cycles, the clips were not fed into the jaws.In order to evaluate the condition of the internal components the device was disassembled; upon disassembling the feeder shoe tab was found to be damaged causing the feeding issues.In addition, 14 clips were found inside the clip track.As part of ethicon¿s quality process, all devices are manufactured, inspected, and released to approved specifications.No conclusion could be reached regarding what may have caused the reported incident.
 
Event Description
It was reported that during a laparoscopic procedure, the clip was not fed into the jaw properly.Another device was used to complete the case.There were no adverse consequences to the patient.No further information is available.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LIGAMAX-5MM ENDO CLIP APPLIER
Type of Device
CLIP, IMPLANTABLE
Manufacturer (Section D)
ETHICON ENDO-SURGERY, LLC.
475 calle c
guaynabo 00969
*  00969
Manufacturer (Section G)
ETHICON ENDO-SURGERY, LLC.
475 calle c
guaynabo 00969
*   00969
Manufacturer Contact
orla o'mahony
475 calle c
guaynabo 
*  
329348013
MDR Report Key15305105
MDR Text Key305530400
Report Number3005075853-2022-05607
Device Sequence Number1
Product Code FZP
UDI-Device Identifier10705036001843
UDI-Public10705036001843
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
K050344
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign
Reporter Occupation Other
Type of Report Initial
Report Date 08/26/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberEL5ML
Device Catalogue NumberEL5ML
Device Lot NumberX94R53
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/03/2022
Initial Date Manufacturer Received 08/03/2022
Initial Date FDA Received08/26/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured04/09/2022
Is the Device Single Use? Yes
Type of Device Usage Initial
Patient Sequence Number1
-
-