It was reported that the procedure was to treat an internal carotid artery.An rx.014 acculink 7-10/40 stent was advanced to the lesion, although resistance with a guide wire was noted it ultimately reached the lesion.Deployment was initiated, however, the stent was released to 1/3, there was great resistance in retracting the handle and the stent could not be released any more.It was decided to withdraw the device and replaced with a non-abbott stent that successfully completed the procedure.There were no adverse patient effects and no clinically significant delay in the procedure.No additional information was provided.
|
The device was not returned for analysis.A review of the lot history record identified no manufacturing nonconformities issued to the reported lot that would have contributed to this event.Additionally, a review of the complaint history identified no other incidents from this lot.As there was no reported damage noted to the device during the inspection prior to use, it is possible that the distal shaft was bent due to the anatomy and/or coagulation of blood and/or contrast on the guide wire resulted in the reported difficult to advance; however this cannot be confirmed.Additionally, it is possible that the distal shaft was bent due to the anatomy and/or interaction with other devices preventing the shaft lumens from moving freely thus resulting in resistance with the slider and difficulty deploying the stent; however, this could not be confirmed.The investigation determined a conclusive cause for the reported difficulties cannot be determined.There is no indication of a product quality issue with respect to manufacture, design or labeling of the device.
|