Model Number COAGUCHEK INRANGE |
Device Problem
Image Display Error/Artifact (1304)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 08/03/2022 |
Event Type
malfunction
|
Manufacturer Narrative
|
The customer's meter was requested for investigation.
|
|
Event Description
|
It was reported that there was an issue with the display of coaguchek vantus meter serial number (b)(4).The meter display was entirely black, but a test could still be performed and the result could be transmitted wirelessly.The batteries were replaced, but the meter display remained entirely black.A display test could not be performed.The meter was not dropped or exposed to liquids.
|
|
Manufacturer Narrative
|
The meter was returned for investigation.The device could be switched on, but the display remained dark.The tone of the beeper and the signal of an error message could be heard.A replacement display showed the same error.The display symbols could be seen on closer inspection of both displays.The backlight of the meter failed due to a defect of the circuit board or electronic component on the circuit board.Medwatch fields d9 and h3 have been updated.
|
|