Device evaluation of the monitor and electrode belt has been completed.The evaluation included review of downloaded software flag files on the day of the event and incoming functional testing.During the incoming functional testing, a 1hz simulated normal sinus rhythm signal was applied to the ecg electrodes, followed by a 5hz simulated treatable arrhythmia signal which verified proper performance of the detection algorithm.During the transition to the 5hz signal, the device was confirmed to properly enter into a treatment sequence which includes a verification of the tactile vibration alarm, audio messaging, and siren alarms, as well as a test of the pulse delivery circuitry.The pulse delivery circuitry test verified proper charging of the high voltage capacitors and proper delivery of five full energy 150j biphasic pulses.The functional testing confirmed proper response button functionality, ecg acquisition, detection algorithm performance, and pulse delivery functionality.There is no indication of a product malfunction.
|
A us distributor contacted zoll to report that a patient passed away on (b)(6) 2022 while reportedly wearing the lifevest.The patient received three appropriate treatments, two which converted the arrhythmias to a slower rhythm and one which resulted in post shock asystole.The patient also received an inappropriate treatment. the device was started up at 17:46:04 on (b)(6) 2022.At 19:22:10, an arrhythmia was detected.Ecg shows vf.At 19:22:46, the patient received the first appropriate treatment.The rhythm at the time of treatment was vf.The post shock rhythm was asystole for 35 seconds transitioning to sinus rhythm @ 60 bpm with pvc.Post-shock asystole is a known and potentially adverse outcome of defibrillation therapy.At 19:23:57, the patient received the second appropriate treatment.The rhythm at the time of treatment was vf.The post shock rhythm was sinus bradycardia @ 20 bpm.At 19:24:29, the patient received an inappropriate treatment.Oversensing of cardiac activity contributed to the false detection.The rhythm at the time of treatment was sinus rhythm @ 70 bpm.The post shock rhythm was sinus rhythm @ 70 bpm with pac.At 19:46:07, an arrhythmia was detected.Ecg shows vt @ 290 bpm.At 19:46:39, the patient received the third appropriate treatment.The rhythm at the time of treatment was vf.The post shock rhythm was sinus bradycardia @ 30 bpm.The device was shut down at 20:29:34 on (b)(6) 2022.
|