Catalog Number CON-HL-390 |
Device Problem
Mechanical Problem (1384)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 10/07/2019 |
Event Type
malfunction
|
Manufacturer Narrative
|
The manufacturing (dhr) device history record review was not performed because the original (dhr) is no longer applicable as the device was manufactured in 1996.No causes or potential causes of the customer's reported problem were found during the review of service and repair records.The product device was received for evaluation.Visual and functional testing were performed.Visual inspection found the enclosure is dirty and scuffed.The water tank cover is missing seal and reservoir is contaminated.Main block is dirty, line cord is old and worn.Functional testing found was able to duplicate reported problem.The root cause of the reported issue was found to be a broken micro switch.Device information is unknown.No product information has been provided to date.Operator of device is unknown.No product information has been provided to date this remediation mdr was generated under protocol (b)(4), as a result of warning letter cms# (b)(4).
|
|
Event Description
|
It was reported that there is a defective micro switch.No patient injury was reported.
|
|