• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ST PAUL LEVEL 1 HOTLINE FLUID WARMER, 390; WARMER, THERMAL, INFUSION FLUID

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ST PAUL LEVEL 1 HOTLINE FLUID WARMER, 390; WARMER, THERMAL, INFUSION FLUID Back to Search Results
Catalog Number CON-HL-390
Device Problem Mechanical Problem (1384)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 10/07/2019
Event Type  malfunction  
Manufacturer Narrative
The manufacturing (dhr) device history record review was not performed because the original (dhr) is no longer applicable as the device was manufactured in 1996.No causes or potential causes of the customer's reported problem were found during the review of service and repair records.The product device was received for evaluation.Visual and functional testing were performed.Visual inspection found the enclosure is dirty and scuffed.The water tank cover is missing seal and reservoir is contaminated.Main block is dirty, line cord is old and worn.Functional testing found was able to duplicate reported problem.The root cause of the reported issue was found to be a broken micro switch.Device information is unknown.No product information has been provided to date.Operator of device is unknown.No product information has been provided to date this remediation mdr was generated under protocol (b)(4), as a result of warning letter cms# (b)(4).
 
Event Description
It was reported that there is a defective micro switch.No patient injury was reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LEVEL 1 HOTLINE FLUID WARMER, 390
Type of Device
WARMER, THERMAL, INFUSION FLUID
Manufacturer (Section D)
ST PAUL
1265 grey fox rd.
st. paul MN 55112
Manufacturer (Section G)
ST PAUL
1265 grey fox rd.
st. paul MN 55112
Manufacturer Contact
jim vegel
6000 nathan lane north
minneapolis, MN 55442
MDR Report Key15314925
MDR Text Key305268803
Report Number3012307300-2022-16569
Device Sequence Number1
Product Code LGZ
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K911383
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type User Facility
Reporter Occupation Other
Type of Report Initial
Report Date 08/29/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received08/29/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Catalogue NumberCON-HL-390
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/20/2019
Was the Report Sent to FDA? No
Date Manufacturer Received04/08/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured07/23/2010
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-