No lot number was provided; therefore, device history record review could not be completed.A product sample was received for evaluation.Visual and functional testing were performed.As a result of observing the sample, it was confirmed that the catheter was broken and the catheter around the broken position was extended.When observed with a microscope, it was confirmed that the fracture surface was crushed into an elliptical shape, which closely resembled the trace of fracture caused by being sandwiched between sharp blades such as scissors.The root cause of the reported issue was found to be highly probable that the catheter was damaged due to contact with scissors during use.Udi is unknown.No product information has been provided to date.This remediation mdr was generated under protocol b10009406, as a result of warning letter cms# (b)(4).
|