It was reported that during preparation of an emboshield nav6 embolic protection system (eps), the filtration element was loaded into the delivery catheter pod (dc pod) and flushed, however, when removing the nav6 from the tray, resistance was met, causing the tip coil of the barewire to unravel and detach.The nav6 was not used.There was no patient involvement and no clinically significant delay in the procedure.A new same size nav6 was used to successfully complete the procedure no additional information was provided.
|
Visual analysis was performed on the returned device.The reported difficulty removing was unable to be confirmed as the tray was not returned.The stretched tip coils and break were confirmed.A review of the lot history record identified no manufacturing nonconformities issued to the reported lot that would have contributed to this event.Additionally, a review of the complaint history identified no other similar complaints.The investigation determined that the reported damage to the barewire was likely related to difficulties during removal from the packing.Based on the reported information, it is likely that after the filter was successfully loaded into the delivery catheter, the barewire tip became caught on the external loading funnel or dispenser hoop resulting in resistance, stretching of the tip coils, and causing the core to break, but remain held together by the tip coils.There is no indication of a product quality issue with respect to the design, manufacture, or labeling of the device.
|