Model Number HL-390
Device Problem Failure to Calibrate (2440)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 02/07/2020
Event Type  malfunction  
Manufacturer Narrative
A device history record (dhr) review was conducted which indicated all inspection were completed and no issues were noted during manufacture.The suspect device was returned for investigation and a visual and functional testing was performed.Physical evaluation shows wear and tear damage in the enclosure, tank cover, front cover, and line cord.The root cause to the suspect device to fail electrical testing is to faulty line cord was found.The root cause to this component failure was unable to determine.Action taken was to replace the enclosure, tank cover, front cover and the line cord; performed preventive maintenance check and testing passed.This remediation mdr was generated under protocol b10009406, as a result of warning letter cms# (b)(4).
 
Event Description
It was reported that the suspect device unable to pass the electrical safety testing: no patient injury was reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LEVEL 1 HOTLINE 3 BLOOD AND FLUID WARMER
Type of Device
WARMER, THERMAL, INFUSION FLUID
Manufacturer (Section D)
ST PAUL
1265 grey fox rd.
st. paul MN 55112
Manufacturer (Section G)
ST PAUL
1265 grey fox rd.
st. paul MN 55112
Manufacturer Contact
jim vegel
6000 nathan lane north
6720 bertner ave
minneapolis, MN 55442
MDR Report Key15319317
MDR Text Key305156486
Report Number3012307300-2022-16665
Device Sequence Number1
Product Code LGZ
UDI-Device Identifier10695085002796
UDI-Public10695085002796
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K911383
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type User Facility
Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date 08/30/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received08/30/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberHL-390
Device Catalogue NumberHL-390
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/15/2020
Was the Report Sent to FDA? No
Date Manufacturer Received04/12/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured03/17/2016
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-