MAUDE Adverse Event Report: ROCHE DIAGNOSTICS COAGUCHEK ® XS SYSTEM; PROTHROMBIN TIME MONITOR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ROCHE DIAGNOSTICS COAGUCHEK ® XS SYSTEM; PROTHROMBIN TIME MONITOR Back to Search Results
Model Number XS PST
Device Problem Image Display Error/Artifact (1304)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 08/04/2022
Event Type  malfunction  
Event Description
There was an allegation of a display issue for a coaguchek xs meter.The device's display has missing segments.During a display check, the "8's" in the middle of the screen were not complete and there were several segments missing.No misinterpretation of a result was reported.
 
Manufacturer Narrative
Occupation is patient/consumer.The device was requested to be returned for investigation.The investigation is ongoing.
 
Manufacturer Narrative
Sections d9 and h3 were updated to reflect returned product.The meter was returned.The investigation found that the device's battery compartment is contaminated by a leaked battery and the circuit board is contaminated by penetrated liquid which caused the alleged issue.Product labeling states: "it is important to follow the procedures below to clean and disinfect the meter.Failure to follow these procedures may cause malfunction of the meter.Do not use sprays of any sort; ensure that swab or cloth is only damp, not wet" and "wipe away residual moisture and fluids after cleaning the housing." the investigation determined the issue to be caused by improper handling or maintenance.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
COAGUCHEK ® XS SYSTEM
Type of Device
PROTHROMBIN TIME MONITOR
Manufacturer (Section D)
ROCHE DIAGNOSTICS
9115 hague road
indianapolis IN 46250 0457
Manufacturer (Section G)
ROCHE DIAGNOSTICS GMBH
sandhoferstrasse 116
na
mannheim (baden-wurttemberg) 68305
GM   68305
Manufacturer Contact
amy nelson
9115 hague road
na
indianapolis, IN 46250
3174767531
MDR Report Key15321138
MDR Text Key305492649
Report Number1823260-2022-02638
Device Sequence Number1
Product Code GJS
UDI-Device Identifier00365702125100
UDI-Public00365702125100
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K062925
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Consumer
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 10/13/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received08/30/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Model NumberXS PST
Device Catalogue Number04837738001
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Date Manufacturer Received09/19/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-