Model Number XS PST |
Device Problem
Image Display Error/Artifact (1304)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 08/04/2022 |
Event Type
malfunction
|
Event Description
|
There was an allegation of a display issue for a coaguchek xs meter.The device's display has missing segments.During a display check, the "8's" in the middle of the screen were not complete and there were several segments missing.No misinterpretation of a result was reported.
|
|
Manufacturer Narrative
|
Occupation is patient/consumer.The device was requested to be returned for investigation.The investigation is ongoing.
|
|
Manufacturer Narrative
|
Sections d9 and h3 were updated to reflect returned product.The meter was returned.The investigation found that the device's battery compartment is contaminated by a leaked battery and the circuit board is contaminated by penetrated liquid which caused the alleged issue.Product labeling states: "it is important to follow the procedures below to clean and disinfect the meter.Failure to follow these procedures may cause malfunction of the meter.Do not use sprays of any sort; ensure that swab or cloth is only damp, not wet" and "wipe away residual moisture and fluids after cleaning the housing." the investigation determined the issue to be caused by improper handling or maintenance.
|
|