Catalog Number HL-90
Device Problem Defective Component (2292)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 01/01/1901
Event Type  malfunction  
Manufacturer Narrative
A product sample was received for evaluation.Visual and functional testing were performed.Visual inspection revealed the pole clamp was damaged and it was an old model.The reported issue was confirmed.No action was taken due to the age and condition of the device.The root cause of the reported issue was found to be user interface and age of the device.A device history record (dhr) review was conducted which indicated all inspections were completed and no issues were noted during manufacture.No information has been provided to date.This remediation mdr was generated under protocol b10009406, as a result of warning letter cms# 617147.
 
Event Description
It was reported that there was the plastic around the pole clamp at the end of the screw is now gone and it will not screw to the iv pole.No patient injury was reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HOTLINE BLOOD AND FLUID WARMER
Type of Device
WARMER, THERMAL, INFUSION FLUID
Manufacturer (Section D)
OAKDALE
3350 granada ave n
oakdale MN 55128
Manufacturer (Section G)
OAKDALE
3350 granada ave n
oakdale MN 55128
Manufacturer Contact
jim vegel
6000 nathan lane north
receiving dock
minneapolis, MN 55442
MDR Report Key15321547
MDR Text Key305361136
Report Number3012307300-2022-16793
Device Sequence Number1
Product Code LGZ
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K001764
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type User Facility
Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date 08/30/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received08/30/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Catalogue NumberHL-90
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/16/2020
Was the Report Sent to FDA? No
Date Manufacturer Received04/12/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured04/28/1995
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-