Catalog Number 4590PG
Device Problems Obstruction of Flow (2423); Defective Device (2588)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Type  malfunction  
Event Description
It was reported that the suspect device was defective due to its inability to aspirate: no patient injury was reported.
 
Manufacturer Narrative
A device history record (dhr) review was conducted which indicated all inspection were completed and no issues were noted during manufacture.The original suspect device was not received in the manufacturing for evaluation; however, an unopened original package was received sent by the customer.The returned sample was evaluated and preformed functional testing in several different scenarios which the returned sample passed all functional testing.The issue that was reported by the customer could not be duplicated with the returned sample, therefore, the complaint could not be confirmed.Will continue to monitor any new information and will take actions accordingly.No problem found based on the provided sample.Udi # is unknown.No information is provided to date.This remediation mdr was generated under protocol (b)(4), as a result of warning letter cms# (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PORTEX ARTERIAL BLOOD SAMPLING KITS
Type of Device
TUBES, VIALS, SYSTEMS, SERUM SEPARATORS, BLOOD COL
Manufacturer (Section D)
SMITHS MEDICAL ASD, INC.
10 bowman dr.
keene NH 03431
Manufacturer (Section G)
SMITHS MEDICAL ASD, INC.
10 bowman dr.
keene NH 03431
Manufacturer Contact
jim vegel
6000 nathan lane north
minneapolis, MN 55442
MDR Report Key15321926
MDR Text Key305424685
Report Number3012307300-2022-16802
Device Sequence Number1
Product Code JKA
Combination Product (y/n)N
Reporter Country CodeFR
PMA/PMN Number
K952516
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional
Reporter Occupation Pharmacist
Type of Report Initial
Report Date 08/30/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received08/30/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date04/13/2023
Device Catalogue Number4590PG
Device Lot Number3968638
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/07/2020
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received04/14/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured05/04/2020
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient SexFemale
-
-