SMITHS MEDICAL ASD, INC. PORTEX ARTERIAL BLOOD SAMPLING KITS; TUBES, VIALS, SYSTEMS, SERUM SEPARATORS, BLOOD COL
|
Back to Search Results |
|
Catalog Number 4590PG |
Device Problems
Obstruction of Flow (2423); Defective Device (2588)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Type
malfunction
|
Event Description
|
It was reported that the suspect device was defective due to its inability to aspirate: no patient injury was reported.
|
|
Manufacturer Narrative
|
A device history record (dhr) review was conducted which indicated all inspection were completed and no issues were noted during manufacture.The original suspect device was not received in the manufacturing for evaluation; however, an unopened original package was received sent by the customer.The returned sample was evaluated and preformed functional testing in several different scenarios which the returned sample passed all functional testing.The issue that was reported by the customer could not be duplicated with the returned sample, therefore, the complaint could not be confirmed.Will continue to monitor any new information and will take actions accordingly.No problem found based on the provided sample.Udi # is unknown.No information is provided to date.This remediation mdr was generated under protocol (b)(4), as a result of warning letter cms# (b)(4).
|
|